Rapid Point-of-Care Assay of Enoxaparin Anticoagulant Efficacy in Whole Blood
Published on: October 12, 2012
Precision of INR measured with a patient operated whole blood coagulometer
Jørn Attermann1, Niels T Andersen, Helle Korsgaard
1Department of Biostatistics, University of Aarhus, Vennelyst Boulevard 6, DK-8000 Aarhus C, Denmark. jorn.attermann@bisam.dk
Introduction:
The objective of the present study was to evaluate the precision of a portable whole blood coagulometer (CoaguChek S) in the hands of self-managing patients on oral anticoagulant therapy (OAT).
Materials And Methods:
Fifteen patients on self-managed OAT performed measurements of INR by two types of portable whole blood coagulometers (CoaguChek and CoaguChek S) at home for 10 weeks.
Results:
The coefficient of variation (CV) of INRs determined at home by CoaguChek S by patients on self-managed OAT was 5.5% (95% confidence limits: 4.9%, 6.1%). The biological CV of INR within and between patients was 15.0% and 14.7%, respectively.
Conclusion:
The precision of CoaguChek S is satisfactory.
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.