Recovery and biological activity of filgrastim after injection through silicone rubber catheters

J M McCullough1, E Sprentall-Nankervis, C A Potcova

  • 1Department of Pharmacy, Botsford General Hospital, 28050 Grand River Avenue, Farmington Hills, MI 48336, USA.

Insights

Filgrastim recovery through silicone catheters showed a drug loss with smaller injection volumes. However, the recovered filgrastim maintained its biological activity, with improved recovery after a second injection.

Area of Science:

  • Pharmacology
  • Biotechnology
  • Medical Devices

Background:

  • Silicone rubber catheters are commonly used for drug delivery.
  • Ensuring drug integrity and efficacy after administration through medical devices is crucial.
  • Filgrastim is a vital biologic for stimulating neutrophil production.

Purpose of the Study:

  • To evaluate the recovery and biological activity of filgrastim after injection through silicone rubber catheters.
  • To determine the impact of injection volume and repeated use on filgrastim recovery.
  • To assess if filgrastim retains its biological function post-catheter administration.

Main Methods:

  • Filgrastim solutions of varying volumes were injected through silicone catheters.
  • Drug concentrations in expelled fluid were measured using enzyme-linked immunosorbent assay (ELISA).
  • Biological activity was assessed via a mitogenic cell-culture assay.

Main Results:

  • A loss of filgrastim (≥10%) occurred with smaller injection volumes (0.17, 0.34, 0.68 mL) after the first use.
  • Drug recovery improved after a second injection through the same catheter, exceeding 90% for most volumes.
  • Recovered filgrastim demonstrated no loss in biological activity.

Conclusions:

  • Silicone rubber catheters can lead to filgrastim loss, particularly with smaller volumes.
  • Repeated use of the same catheter improves filgrastim recovery.
  • The biological activity of filgrastim is preserved following administration through silicone catheters.