Evaluation of two new high-sensitivity methods for C-reactive protein
Snjezana Rothkrantz-Kos1, Otto Bekers, Armand Gubbels
1Department of Clinical Chemistry, University Hospital Maastricht, 6202 AZ Maastricht, The Netherlands.
Annals of Clinical Biochemistry
|July 26, 2003
Summary
The BN ProSpec hs-CRP assay demonstrated superior performance compared to the Synchron LX 20 PRO hs-CRP method. Further standardization is needed for both high-sensitivity C-reactive protein assays across all concentration ranges.
Area of Science:
- Clinical Chemistry
- Biomarker Assay Development
- Laboratory Medicine
Background:
- High-sensitivity C-reactive protein (hs-CRP) assays are crucial for routine laboratory diagnostics.
- A single, reliable hs-CRP method across all concentration ranges is needed for practical implementation.
- Two novel hs-CRP assays were evaluated for their performance in both low and high concentration ranges.
Purpose of the Study:
- To evaluate the performance of two new hs-CRP assays: BN ProSpec and Synchron LX 20 PRO.
- To compare these new assays against an existing hs-CRP IMMAGE method and a routine Synchron LX 20 CRP method.
- To assess assay linearity, precision, and susceptibility to sample turbidity.
Main Methods:
- Comparative analytical evaluation of BN ProSpec hs-CRP and Synchron LX 20 PRO hs-CRP assays.
- Reference method: hs-CRP IMMAGE assay; high concentration comparison: routine Synchron LX 20 CRP assay.
- Analysis of 521 patient samples, estimation of reference values in 291 blood donors, and turbidity influence assessment.
Main Results:
- BN ProSpec demonstrated linearity down to 0.2 mg/L with precision (CV) < or =3.7% across the 0.2-250 mg/L range.
- Synchron LX 20 PRO showed some linearity discrepancies and precision (CV) < or =6.1%.
- Significant discrepancies between methods were noted, particularly in high concentration ranges; BN ProSpec outperformed Synchron LX 20 PRO regarding precision and turbidity influence.
Conclusions:
- The BN ProSpec hs-CRP method exhibited superior overall performance compared to the Synchron LX 20 PRO hs-CRP method.
- While Synchron LX 20 PRO showed acceptable performance, standardization across hs-CRP assays requires significant improvement.
- Further efforts are necessary to achieve assay standardization for both low and high concentration ranges in hs-CRP testing.