Related Experiment Videos
Automated large-volume sample stacking procedure to detect labeled peptides at picomolar concentration using
Nathalie Siri1, Pierre Riolet, Christophe Bayle
1Picometrics, 10 Avenue de l'Europe, 31520 Ramonville, France.
Abstract:
We have developed an automated large-volume sample stacking (LVSS) procedure to detect fluorescein isothiocyanate-labeled peptides in the picomolar range. The injection duration is 10 min at 50 mbar to fill 62% of the capillary volume to the detection cell. The calculated limit of detection (S/N=3), filling 1% of the capillary volume, is 74 pM for bradykinin and 45 pM for L-enkephalin with samples diluted in water and analyzed in a 50 mM borate buffer, pH 9.2. With the automated LVSS system, the limits of detection are 7 pM for bradykinin, 3 pM for L-enkephalin and 2 pM for substance P. LVSS is shown to be quantitative from 500 to 10 pM.