Related Experiment Video
Updated: Aug 18, 2026

Stereotactic Radiosurgery for Gynecologic Cancer
Published on: April 17, 2012
Comprehensive quality assurance for stereotactic radiosurgery treatments
1Radiation Physics Department, Princess Margaret Hospital, University Health Network, 610 University Avenue, Toronto, ON, M5G 2M9, Canada.
Abstract:
We have used a commercially available high precision Lucy phantom to perform comprehensive quality assurance for stereotactic radiosurgery treatments. The quantitative evaluation of system uncertainties included imaging, planning and treatment delivery systems. The quality assurance tests showed that the well-defined targets were identified to within +/-1 mm in all the imaging modalities. The pre-known target volumes were reproduced within 2 cm3 in both MR and CT. The planned target was delivered within 2% of the prescribed dose and to within 2 mm accuracy. The inaccuracy in the isocentre position at the Linac was less than 1.2 mm. The maximum error observed in the depth helmet was 0.5 mm and the overall uncertainty was within 0.23 mm. We have also established a quality assurance program based on the study and proposed the tolerance and the frequency of the tests required to be carried out. The tests were carried out using a Radionics planning system and delivered on a Varian Clinac 2100 linear accelerator machine. These tests also established a base line for future comparisons.

