Related Experiment Video
Updated: Sep 20, 2026

A Tailored HPLC Purification Protocol That Yields High-purity Amyloid Beta 42 and Amyloid Beta 40 Peptides, Capable of Oligomer Formation
Published on: March 27, 2017
Beta2-microglobulin-selective adsorbent column (Lixelle) for the treatment of dialysis-related amyloidosis
Kazuo Suzuki1, Masami Shimazaki, Hidetoshi Kutsuki
1Medical Devices Division, Kaneka Corporation, Osaka, Japan. kazuo.suzuki@kn.kaneka.co.jp
Abstract:
Lixelle is a direct hemoperfusion-type adsorbent column developed for selective elimination of beta2-microglobulin (beta2-m) from the circulating blood of patients with dialysis-related amyloidosis (DRA). Lixelle S-35, that has a column volume of 350 mL, efficiently eliminates beta2-m and improves symptoms of DRA such as joint pain and nocturnal awakening. The performance of Lixelle S-15, that has a column volume of 150 mL, is dependent on the dialysis membrane used in combination with the adsorbent column in a hemodialysis circuit. The combination of S-15 and a dialysis membrane with a high beta2-m clearance eliminates an amount of beta2-m that is nearly equal to the amount removed by using S-35. Treatment with S-15 for 6 months improved joint pain with an efficacy similar to that observed when using S-35. The major adverse effects were hypotension and decrease in hematocrit, and these incidences were much less in S-15 compared to S-35.
Related Concept Videos
Dialysis
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis