Effect of pore size on plate height for 23-mer oligonucleotides separated using fully porous particles and
Daniel Meston1, Nolan Eckmann1, Ronard Pabi1
1Department of Chemistry, Gustavus Adolphus College, Saint Peter, MN, 56082, United States.
Abstract:
Identification and quantification of impurities in therapeutic oligonucleotides (ONs) by liquid chromatography is a critical component of control strategies used in pharmaceutical manufacturing. While ion-pairing reversed-phase liquid chromatography (IPRP-LC) remains the gold standard for these separations, the fundamental influence of stationary phase physical parameters on the kinetic performance of ON separations remains under-characterized compared to the knowledge base that exists for small molecule separations. In this work we have investigated the effect of stationary phase pore size and particle size on plate height for a 23-mer surrogate therapeutic oligonucleotide under IPRP conditions. Plate heights (H) were determined under isocratic conditions for eight fully porous stationary phases, including polymeric and silica-based materials, with pore sizes ranging from 80 to 4000 Å and particle sizes from 1.8 to 5.0 µm. To enable comparability across different materials, mobile phase conditions were carefully adjusted to obtain plate height data at a constant retention factor of 2. The results show that intraparticle pore size is a stronger determinant of plate height than particle size for the ON studied here. For silica-based materials, increasing the pore size from 80 Å to 300 Å resulted in significantly lower plate heights and a weaker dependence of efficiency on mobile phase velocity. Remarkably, a column packed with 3.5 µm, 300 Å particles yielded a three-fold lower plate height than a 1.8 µm, 80 Å column at practically relevant flow rates. Moreover, we found that the observed trends in isocratic plate heights translated well to solvent gradient elution conditions that are more relevant to practical separations of ON mixtures. Finally, results from a brief set of pore penetration experiments using poly-thymidine standards suggest that diffusion of ONs larger than five bases through the intraparticle pores of conventional stationary phase materials (80-120 Å) is probably highly hindered. These findings underscore the importance of using wider-pore (e.g., >200 Å) stationary phases to maximize resolution in the analysis of therapeutic ONs.
Related Concept Videos
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Analyte Adsorption and Distribution
High-Performance Liquid Chromatography: Elution Process