High-resolution backscatter electron imaging of colloidal gold in LVSEM
Related Experiment Video
Updated: Sep 20, 2026

Scalable Isolation and Purification of Extracellular Vesicles from Escherichia coli and Other Bacteria
Published on: October 13, 2021
S Erlandsen1, Y Chen, C Frethem
1Department of Genetics, Cell Biology and Development, University of Minnesota Medical School, Minneapolis, MN 55455, USA. erlan001@umn.edu
Abstract:
High-resolution backscatter electron (BSE) imaging of colloidal gold can be accomplished at low voltage using in-lens or below-the-lens FESEMs equipped with either Autrata-modified yttrium aluminium garnet (YAG) scintillators doped with cerium, or with BSE to secondary electron (SE) conversion plates. The threshold for BSE detection of colloidal gold was 1.8 keV for the YAG detector, and the BSE/SE conversion was sensitive down to 1 keV. Gold particles (6, 12 and 18 nm) have an atomic number of 79 and were clearly distinguished at 500,000x by materials contrast and easily discriminated from cell surfaces coated with platinum with an atomic number of 78. BSE imaging was relatively insensitive to charging, and build up of carbon contamination on the specimen was transparent to the higher energy BSE.
Related Concept Videos
Immunogold Electron Microscopy
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Preparation of Samples for Electron Microscopy
Overview of Electron Microscopy
Super-resolution Fluorescence Microscopy