First clinical experiences with a new cervical fixation device - technical report.
1Neurosurgical Department, University of Ulm/Günzburg, Germany. wolfgang.boerm@medizin.uni-ulm.de
Zentralblatt Fur Neurochirurgie
|September 17, 2003
Summary
The StarLock cervical fixation device offers improved screw pullout resistance and simplified rod insertion, reducing operation time. This system facilitates easier implantation and computer-navigated screw placement in cervical spine surgeries.
Area of Science:
- Orthopedic surgery
- Spinal instrumentation
Background:
- Cervical spine fixation requires reliable and easy-to-use instrumentation.
- Existing systems may present challenges in rod insertion and screw placement.
Purpose of the Study:
- To evaluate the initial clinical outcomes of the novel StarLock cervical fixation device.
- To assess the ease of implantation and effectiveness of the StarLock system in achieving spinal stabilization.
Main Methods:
- A clinical series of ten patients undergoing cervical fixation surgery.
- Implantation of the StarLock device, including insertion of 62 lateral mass screws.
- Radiological assessment of screw placement and spinal stabilization.
Main Results:
- Successful radiologically proven stabilization was achieved in most cases.
- The device demonstrated high angular tolerance, simplifying rod insertion and reducing operation time.
- A low rate of insufficient screw placement (3.2%) was observed, attributed to surgical difficulties unrelated to the device.
Conclusions:
- The StarLock cervical fixation device is a user-friendly system that simplifies rod application, particularly for longer constructs.
- Its design facilitates computer-navigated screw placement and offers high pullout resistance.