[Genotype reliability of short tandem repeats typing from minute DNA]
De-jian Lu1, Hong-yu Sun, Li-xian Chen
1Department of Forensic Biology, Zhongshang Medical College, Guangzhou 510089, China. dejian6182@sina.com
Objective:
To understand the forensic practice of DNA profiling from minimal amount of DNA.
Methods:
Serial dilutions of DNA were amplified with the PowerPlex 16 System Kit, then the genotyping of short tandem repeat(STR) was performed by ABI 377 DNA automated Sequencer.
Results:
When the mount of DNA template was less than 250 pg, allelic drop-out apparently occurred at several loci. Other disturbed peaks, such as artefact bands and imbalanced heterozygote, also presented.
Conclusion:
The anomalous results may result in incorrect genotyping. Careful and comprehensive considerations are needed to interpret the STR profile of minute DNA.