Related Experiment Video
Updated: Aug 13, 2026

Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Turbulent airflow meter for long-term monitoring in patient-ventilator circuits
Abstract:
A new type of flowmeter is described which operates on the principle that pressure drop (deltaP) produced by turbulent volume flow (V) through a simple resistance chamber obeys a relation of V = K (deltaP)0.50. Data are given showing that the device follows a true power law with a single exponent for airflows ranging from 3.5 to 270 l/min. Standard instrumentation and a linearization circuit are used with the new flowmeter to provide linear steady-state and alternating airflow measurements up to peak rates of 150 l/min and frequency to 15 Hz. Data are presented comparing integrated airflow readings from the new turbulent flowmeter and a laminar-type flowmeter. The turbulent airflow meter appears to offer increased reliability for long-term patient monitoring use in patient-ventilator circuits.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pipe Flowrate Measurement
The orifice meter is a simple,...