Should we judge a mask by its cover?
1Rayne Laboratory, Department of Medicine (RIE), University of Edinburgh, City Hospital.
Thorax
|July 1, 1992
Summary
Oxygen masks may deliver higher inspired oxygen concentrations than labeled, especially without Venturi attachments. Accurate low oxygen delivery is crucial for emergency use, requiring careful mask selection.
Area of Science:
- Respiratory Medicine
- Biomedical Engineering
Background:
- Oxygen mask packaging often provides tables for inspired oxygen delivery at specific flow rates.
- A new multipurpose oxygen mask designed for use with or without Venturi attachments was evaluated for emergency applications.
Purpose of the Study:
- To assess the accuracy of inspired oxygen concentration delivered by a new multipurpose oxygen mask.
- To compare delivered oxygen concentrations with manufacturer-stated values and evaluate the effect of Venturi attachments.
Main Methods:
- Inspired oxygen concentrations were measured in 12 healthy subjects at rest and 26 breaths/min using a mass spectrometer.
- Oxygen and carbon dioxide levels were recorded at the lips, and inspired oxygen was calculated using the alveolar gas equation.
- Measurements were taken with the mask alone and with 24% and 28% Venturi attachments at various oxygen flow rates.
Main Results:
- Without Venturi attachments, the mask delivered 8-10% higher oxygen concentrations than labeled at 2, 4, and 6 L/min.
- Some individuals received up to 30% more oxygen than expected without the Venturi attachment.
- With Venturi attachments, delivered concentrations were within 2% of the intended 24% and 28%.
Conclusions:
- Labeling on oxygen masks may be inaccurate, potentially leading to incorrect oxygen therapy, especially by non-experts.
- Careful selection of multipurpose masks is essential for emergency use, particularly when precise low oxygen concentrations are required.