Related Experiment Videos
Generation of an osmolality gap-ethanol nomogram from routine laboratory data
1Department of Emergency Medicine, Texas Tech University School of Medicine, El Paso.
Annals of Emergency Medicine
|November 1, 1992
Summary
Laboratories can create custom nomograms for osmolality gap and serum ethanol concentrations. This procedure accounts for specific laboratory formulas and methods, improving accuracy beyond published guidelines.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Toxicology
Background:
- The osmolality gap is a critical indicator in diagnosing conditions like alcohol intoxication.
- Existing nomograms for osmolality gap and serum ethanol may not align with diverse laboratory methodologies.
- Interlaboratory analytical bias and formula selection impact the accuracy of published nomograms.
Purpose of the Study:
- To develop a practical, adaptable procedure for generating laboratory-specific nomograms.
- To describe the relationship between osmolality gap and serum ethanol concentrations.
- To ensure nomogram relevance across different analytical methods and formulas.
Main Methods:
- Retrospective analysis of patient data including serum sodium, glucose, urea nitrogen, ethanol, and osmolality.
- Utilized spreadsheet software for nomogram generation.
- Employed regression analysis and truncated normal distributions for nomogram refinement.
Main Results:
- Published nomograms and common criteria showed significant inconsistencies with laboratory data.
- A novel nomogram, derived from regression analysis, better represented the collected data.
- The developed nomogram is tailored to specific laboratory formulas and analytical methodologies.
Conclusions:
- Laboratory-specific nomograms are essential due to interlaboratory variability and formula choices.
- The described spreadsheet procedure enables laboratories to create their own nomograms.
- This method avoids the need for measuring volatiles other than ethanol, simplifying the process.