An evaluation of pulse oximeters in dogs, cats and horses
Nora S Matthews1, Sherrie Hartke, John C Allen
1Department of Small Animal Medicine and Surgery, Texas A&M University, College Station, TX 77843-4474, USA. nmatthews@cvm.tamu.edu
Objective:
Evaluation of five pulse oximeters in dogs, cats and horses with sensors placed at five sites and hemoglobin saturation at three plateaus.
Study Design:
Prospective randomized multispecies experimental trial.
Animals:
Five healthy dogs, cats and horses.
Methods:
Animals were anesthetized and instrumented with ECG leads and arterial catheters. Five pulse oximeters (Nellcor Puritan Bennett-395, NPB-190, NPB-290, NPB-40 and Surgi-Vet V3304) with sensors at five sites were studied in a 5 x 5 Latin square design. Ten readings (SpO2) were taken at each of three hemoglobin saturation plateaus (98, 85 and 72%) in each animal. Arterial samples were drawn concurrently and hemoglobin saturation was measured with a co-oximeter. Accuracy of saturation measurements was calculated as the root mean squared difference (RMSD), a composite of bias and precision, for each model tested in each species.
Results:
Accuracy varied widely. In dogs, the RMSD for the NPB-395, NPB-190, NPB-290, NPB-40 and V3304 were 2.7, 2.2, 2.4, 1.7 and 2.7% respectively. Failure to produce readings for the NPB-395, NPB-190, NPB-290, NPB-40 and V3304 were 0, 0, 0.7, 0, and 20%, respectively. The Pearson correlation coefficients for the tongue, toe, ear, lip and prepuce or vulva were 0.95, 0.97, 0.69, 0.87 and 0.95, respectively. In horses, the RMSD for the NPB-395, NPB-190, NPB-290, NPB-40 and V3304 were 3.1, 3.0, 4.7, 3.3 and 2.1%, respectively while rates of failure to produce readings were 10, 21, 0, 17 and 60%, respectively. The Pearson correlation coefficients for the tongue, nostril, ear, lip and prepuce or vulva were 0.98, 0.94, 0.88, 0.93 and 0.94, respectively. In cats, the RMSD for all data for the NPB-395, NPB-190, NPB-290, NPB-40 and V3304 were 5.9, 5.6, 7.9, 7.9 and 10.7%, respectively while failure rates were 0, 0.7, 0, 20 and 32%, respectively. The correlation coefficients for the tongue, rear paw, ear, lip and front paw were 0.54, 0.79,.0.64, 0.49 and 0.57, respectively. For saturations above 90% in cats, the RMSD for the NPB-395, NPB-190, NPB-290, NPB-40 and V3304 were 2.6, 4.4, 4.0, 3.5 and 4.8%, respectively, while failure rates were 0, 1.7, 0, 25 and 43%, respectively.
Conclusions And Clinical Relevance:
Accuracy and failure rates (failure to produce a reading) varied widely from model to model and from species to species. Generally, among the models tested in the clinically relevant range (90-100%) RMSD ranged from 2-5% while failure rates were highest in the V3304.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...