Routine cleaning and autoclaving does not remove protein deposits from reusable laryngeal mask devices
Gay Clery1, Joseph Brimacombe, Tisha Stone
1*Department of Anaesthesia and Intensive Care, Cairns Base Hospital, Cairns, Australia; †James Cook University, Townsville, Queensland, Australia; ‡Department of Anaesthesia and Intensive Care Medicine, Leopold-Franzens University, Innsbruck, Austria; and §Central Sterile Supply Department, Cairns Base Hospital, Cairns, Australia.
Unlabelled:
We tested the hypothesis that routine cleaning and autoclaving does not remove protein deposits from reusable laryngeal mask airways (LMAs). All previously used classic and flexible LMAs from a single hospital were tested. Each LMA was hand-washed in an enzymatic solution for 3 min, machine-washed with a disinfectant for 14 min at up to 85 degrees C; dried for 30 min at 75 degrees C, packaged in porous film, and autoclaved for 4 min at 134 degrees C. Protein deposits were detected by immersion in a 1.2% erythrosin B solution, which stains for proteins present on surfaces. The outer surface, inner surface, and edges of the cuff, and the outer and inner surface of the airway tube were examined for the uptake of stain. The severity of staining was scored according to the percentage of surface area stained: nil, mild (>0%-20%), moderate (>20%-80%), and severe (>80%-100%). Fifty-nine LMAs were collected (16 classic and 43 flexible). Staining was detected on 15 (94%) classic LMAs and 39 (91%) flexible LMAs. All staining was graded as mild. Staining was less common on the inner surface than the outer surface or edges for both devices (all P < 0.004). Staining was less common with the airway tube than any location on the cuff (all P < 0.02). Two unused LMAs used as controls were without staining. We conclude that routine cleaning and autoclaving does not remove protein material from reusable LMAs.
Implications:
Routine cleaning and autoclaving does not remove protein material from reusable laryngeal mask airway devices.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Detergent Purification of Membrane Proteins
Physical Methods for Controlling Microbial Growth: Radiation and Filtration