Background:
The clinical efficacy of ICD therapy depends on accurate sensing of intracardiac signals and sensing algorithms. We investigated the occurrence of sensing abnormalities in patients with dual chamber ICDs.
Methods:
The study group consisted of all patients with dual chamber ICDs enrolled in the LESS trial and patients implanted with dual chamber ICDs at a single center between January 1997 and July 2000. Electrograms of spontaneous ventricular arrhythmias requiring device intervention were analyzed.
Results:
A total of 48 patients met the criteria for enrollment. Among the 244 episodes, 215 (88%) were due to ventricular tachycardia and 29 (12%) were due to ventricular fibrillation. Overall undersensing was infrequent with 12 (20%) patients exhibiting on average 2.2 undersensed beats during 26 episodes of ventricular arrhythmias. There was no delay in therapy due to undersensing. Oversensing occurred in 5 (10%) patients resulting in 13 (2.7%) episodes of inappropriate therapy. None of the patients had any lead abnormalities and oversensing resolved after device reprogramming in 4 patients while 1 patient required a separate rate sensing lead. Among patients with oversensing, 4 out of 5 were pacing before the index event while among patients with no oversensing only 5 out of 42 were pacing (P<0.001).
Conclusions:
Dual chamber ICDs demonstrate outstanding accuracy of sensing. However, because of the selection of patient population requiring more frequent pacing, oversensing occurs with a significant frequency. Meticulous evaluation in such patients is necessary to minimize the likelihood of oversensing and inappropriate shocks.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mechanism of Cardiac Arrhythmias