Effect of angular error on tissue Doppler velocities and strain.
Camilla Storaa1, Peter Aberg, Britta Lind
1Division of Medical Engineering, Karolinska Institutet, Huddinge, Sweden. camilla.storaa@labmed.ki.se
Echocardiography (Mount Kisco, N.Y.)
|October 11, 2003
Summary
Ultrasound Doppler
Area of Science:
- Cardiovascular Imaging
- Echocardiography
- Medical Ultrasound
Background:
- Ultrasound Doppler's angle dependency limits velocity measurements to components along the transducer's line of sight.
- This angular error is a significant concern in clinical applications of Doppler echocardiography.
Purpose of the Study:
- To evaluate the clinical impact of angular error on myocardial velocity measurements.
- To assess the effect of correcting for angle dependency on longitudinal strain calculations.
Main Methods:
- Apical two- and four-chamber views were acquired from 43 individuals.
- Myocardial velocities were measured, and angular error was corrected by back-projection onto the longitudinal axis.
- Myocardial strain was calculated from both corrected and uncorrected velocities in 12 myocardial segments per individual.
Main Results:
- The difference between corrected and uncorrected strain values was insignificant in 5 segments and minor (decimal range) in 11 segments.
- The largest discrepancies between measured and corrected strain were observed in apical segments.
- Strain calculations demonstrated greater robustness against angular error compared to velocity measurements.
Conclusions:
- Angular error has minimal impact on longitudinal strain calculations in clinical echocardiography.
- Correction for angular error in ultrasound Doppler measurements of longitudinal strain can be safely omitted.