Related Experiment Video
Updated: Aug 11, 2026

Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
Maximum FIO2 in minimum time depending on the kind of resuscitation bag and oxygen flow
Salvador Quintana1, Jesús Martínez Pérez, Manuel Alvarez
1Intensive Care Unit, Hospital Mútua de Terrassa, Pça Dr. Robert 5, 08221 Terrassa Barcelona, Spain. UCI@mutuaterrassa.es
Objective:
To analyze what FIO2 can be reached, and how long it takes using the different autoinflated resuscitation bags and increasing oxygen flows.
Design:
Experimental analysis on the effect of three different models of autoinflated resuscitation bag and increasing oxygen flows in the final FIO2, and time spent to reach it.
Setting:
Laboratory, with a gas analyzer and a lung simulator to measure inspired FIO2.
Interventions:
Simulated cardiopulmonary resuscitation. Three different autoinflated resuscitation bags were studied; A, the classic one with oxygen delivery directly into the bag, without reservoir, B, a new one without the reservoir device; and C, a new one with the reservoir device properly implemented. Increasing oxygen flows were administered until FIO2 stabilized.
Results:
With model A the maximum FIO2 reached was 0.73 in 70 s using a 20 l/min oxygen flow. With model B the maximum FIO2 reached was 0.65 in 90 s using a 20 l/min oxygen flow. The best FIO2 (0.99) was reached using model C in 55 s with 12 l/min oxygen flow. In the three models a high correlation between oxygen flow and FIO2 was found (r>0.8).
Conclusions:
It is mandatory to use model C resuscitation autoinflated bag with 12 l/min of oxygen flow during the resuscitation maneuvers. Using another autoinflated bag model, maximum oxygen flows (i.e., 20 l/min) are needed. The resuscitation autoinflated bags showed less effectiveness when they were not properly assembled.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Cardiopulmonary Resuscitation II: ACLS Airway Management