Biocompatibility of a physiological pressure sensor
Chao Yang1, Chunfeng Zhao, Lester Wold
1Orthopedic Biomechanics Laboratory, Mayo Clinic, 200 First Street, SW, Rochester, MN 55905, USA.
Abstract:
A newly developed fiber optic micropressure sensor was evaluated for biocompatibility using the International Organization for Standardization (ISO) test standard 10993-6. The test material and an inert control (fused silica glass) were tested in New Zealand white rabbits. Four test specimens were implanted in the paravertebral muscles on one side of the spine about 2-5 cm from the mid-line and parallel to the spinal column. Similarly, four control specimens were implanted on the opposite side. The implantation periods were 1, 4, and 12 weeks to ensure a steady state biological tissue response. Four animals were tested at each time period. Macroscopic and microscopic observations were performed to compare the biological reactions between the test and control materials. There was an inflammatory reaction at 1 week which subsided at 4 weeks. There was fibrous tissue growth near the implant that also decreased over time. Most importantly, there was no significant difference in the biological response between the test and control materials. Therefore, we conclude that the pressure microsensor is biocompatible.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pressure Gauges
