Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
[Artifacts in MRT caused by instruments and implants]
Hansjörg Graf1, Ulrike A Lauer, Thomas Klemm
1Sektion für Experimentelle Radiologie, Eberhard-Karls-Universität Tübingen. hansjoerg.graf@med.uni-tuebingen.de
Abstract:
Metallic instruments and implants can cause severe image artifacts in magnetic resonance imaging (MRI). Besides the properties of the materials and the geometrical arrangement of the devices, the applied MRI sequence type and its parameters (echo time, voxel size, read-out bandwidth, orientations of encoding directions, etc.) play also an important role. These interactions are presented in a systematic survey. A detailed description of the basic physical mechanisms underlying the generation of artifacts is also provided.