[Catheter-based rescue of a split pacemaker lead in the pulmonary artery]
A Drögemüller1, H Schultz, J Senges
1Herzzentrum Ludwigshafen, Kardiologie, Bremserstr. 79, 67063 Ludwigshafen, Germany. a.droegemueller@surfeu.de
Abstract:
This case report describes a successful catheter-based rescue of a split pacemaker lead in the pulmonary artery-as an unusual reason for pleural effusion. Hemorrhagic pleural effusion was diagnosed 7 months after pacemaker exchange due to battery depletion and lead replacement due to an increase of the stimulation threshold. The reason for the pleural effusion was a splitting of the inactive electrode lead into shaft and inner fragment which was dislocated into the pulmonary artery and micro-perforated. The dislocated lead was retrieved using biopsy forceps and a gooseneck snare. Finally it was capped and fixed. Even in an anatomically difficult position, catheter-based rescue is feasible. Not in all pacemaker lead types is cutting the adapter of the inactive pacemaker leads recommendable.