Related Experiment Video
Updated: Aug 30, 2026

Diagnosis of Musculus Gastrocnemius Tightness - Key Factors for the Clinical Examination
Published on: July 7, 2016
Assessment of an electronic goniometer designed to measure equinus contracture
Mathieu Assal1, Jane B Shofer, Eric Rohr
1Rehabilitation Research and Development, Center of Excellence in Limb Loss Prevention and Prosthetic Engineering, Seattle VA Medical Center, WA 98108, USA. mathieu.assal@hcuge.ch
Abstract:
To achieve more objective and repeatable measurements of equinus contracture, we developed the equinometer, a device that allows the measurement of ankle range of motion under controlled torque conditions. This study assessed its accuracy across different subjects and examiners. Two examiners used the equinometer to measure the angle of ankle dorsiflexion at 15 N x m torque on five subjects. Accounting for variation in measurements because of subjects, examiners, and placement of device, we used linear mixed-effects models. Accounting for the variation because of subject, different placements of the equinometer within each subject and the adjustment for the effects of examiner and trial sequence, the standard deviation was 0.94 degrees, 95% confidence interval (0.79 degrees, 1.13 degrees). An upper standard deviation of 1.36 degrees is felt to be acceptable for clinical investigation.
More Related Videos
07:19Clinical Efficacy of Tissue-Bone Homeostasis Manipulation on Soft Tissue Balance and Function in Knee Osteoarthritis
Published on: June 16, 2026
07:51Reliability of A Vibration-Based Elastography Protocol For Assessing Achilles Tendon Stiffness Across Multiple Joint Angles In Elite Athletes
Published on: June 16, 2026