Discrepancies between sodium concentrations measured by the Kodak Ektachem 700 and by dilutional and direct
J A Stone1, J R Moriguchi, D R Notto
1Department of Laboratory Medicine and Pathology, University of Minnesota Hospital and Clinic, Minneapolis 55455.
Abstract:
We have identified rare (approximately 0.2% of all samples), but clinically significant, discrepancies between serum or plasma sodium concentrations measured with the Kodak Ektachem 700's direct ion-selective electrode (ISE) method and concentrations measured with two other analyzers: the Beckman Synchron CX3's dilutional ISE instrument and the Radiometer KNA2 instrument for sodium-potassium analysis by the direct ISE method. The differences do not appear to be related to any previously identified sources of discrepancy, such as variations in triglycerides, bicarbonate, total protein, albumin, or gamma-globulin, the presence of paraproteins, or interference by benzalkonium chloride from heparinized catheters. They occurred despite the use of Gen 04 reference fluid on the Ektachem. We could not identify any drug or family of drugs that the patients had taken in common and that might influence the results. Until this problem is resolved, Ektachem users should be aware of the potential for discrepancies of > 6 mmol/L in measurements of sodium concentrations.
More Related Videos
Related Concept Videos
Determining the pH of Salt Solutions
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides and...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
EDTA: Indirect and Alkalimetric Titration
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...