Accuracy of automatic tube compensation in new-generation mechanical ventilators
Serge Elsasser1, Josef Guttmann, Reto Stocker
1Department of Internal Medicine, Hospital Langenthal, Switzerland.
Objective:
To compare performance of flow-adapted compensation of endotracheal tube resistance (automatic tube compensation, ATC) between the original ATC system and ATC systems incorporated in commercially available ventilators.
Design:
Bench study.
Setting:
University research laboratory.
Subjects:
The original ATC system, Dräger Evita 2 prototype, Dräger Evita 4, Puritan-Bennett 840.
Interventions:
The four ventilators under investigation were alternatively connected via different sized endotracheal tubes and an artificial trachea to an active lung model. Test conditions consisted of two ventilatory modes (ATC vs. continuous positive airway pressure), three different sized endotracheal tubes (inner diameter 7.0, 8.0, and 9.0 mm), two ventilatory rates (15/min and 30/min), and four levels of positive end-expiratory pressure (0, 5, 10, and 15 cm H2O).
Measurements And Main Results:
Performance of tube compensation was assessed by the amount of tube-related (additional) work of breathing (WOBadd), which was calculated on the basis of pressure gradient across the endotracheal tube. Compared with continuous positive airway pressure, ATC reduced inspiratory WOBadd by 58%, 68%, 50%, and 97% when using the Evita 4, the Evita 2 prototype, the Puritan-Bennett 840, and the original ATC system, respectively. Depending on endotracheal tube diameter and ventilatory pattern, inspiratory WOBadd was 0.12-5.2 J/L with the original ATC system, 1.5-28.9 J/L with the Puritan-Bennett 840, 10.4-21.0 J/L with the Evita 2 prototype, and 10.1-36.1 J/L with the Evita 4 (difference between each ventilator at identical test situations, p <.025). Expiratory WOBadd was reduced by 5%, 26%, 1%, and 70% with the Evita 4, the Evita 2 prototype, the Puritan-Bennett 840, and the original ATC system, respectively. The expiratory WOBadd caused by an endotracheal tube of 7.0 mm inner diameter was 5.5-42.2 J/L at a low ventilatory rate and 19.6-82.3 J/L at a high ventilatory rate. It was lowest with the original ATC system and highest with the Evita 4 ventilator (p <.025).
Conclusions:
Flow-adapted tube compensation by the original ATC system significantly reduced tube-related inspiratory and expiratory work of breathing. The commercially available ATC modes investigated here may be adequate for inspiratory but probably not for expiratory tube compensation.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...