Evaluation of the CELL-DYN 3000 differential.
P J Cornbleet1, D Myrick, S Judkins
1Department of Pathology, Stanford University Medical Center, California 94305.
American Journal of Clinical Pathology
|December 1, 1992
Summary
The CELL-DYN 3000 automated differential provides excellent quantitative white blood cell results. However, its suspect flags show limitations in detecting low percentages of abnormal leukocytes and nucleated red blood cells.
Area of Science:
- Hematology
- Clinical Pathology
- Laboratory Medicine
Background:
- Automated blood cell differentials are crucial for accurate patient diagnosis.
- Evaluating the performance of new hematology analyzers is essential for clinical laboratories.
- The CELL-DYN 3000 is an automated system for complete blood count and differential analysis.
Purpose of the Study:
- To evaluate the performance of the CELL-DYN 3000 automated differential analyzer.
- To assess the accuracy of instrument-generated suspect flags for abnormal leukocytes.
- To determine the reliability of abnormal cell estimates provided by the CELL-DYN 3000.
Main Methods:
- Comparison of CELL-DYN 3000 results with a 400-cell manual differential.
- Evaluation of suspect flags for blasts, variant lymphocytes, nucleated red blood cells (NRBCs), and immature granulocytes (IGs).
- Analysis of sensitivity and specificity for detecting various abnormal leukocyte populations and platelet clumping.
Main Results:
- Excellent quantitative white blood cell differential results were observed.
- Combined specificity for BLAST, VARIANT LYMPH, NRBC, WBC, or DIFF flags was 82.6%.
- Combined sensitivity for detecting abnormal leukocytes or platelet clumping was 81.6%.
- The instrument struggled to flag low percentages (< or = 5%) of nucleated erythrocytes, lymphoid blasts, hairy cells, or >5% reactive lymphocytes.
- Specificity of the immature granulocyte (IG) flag was 94.9% for IG counts >= 3%.
- Sensitivity of the IG flag varied from 41.7% (>= 1% IG) to 100% (>= 3% IG).
- Flagging performance for band neutrophils showed poor sensitivity and specificity.
Conclusions:
- The CELL-DYN 3000 offers reliable quantitative differential analysis.
- Instrument flags require careful review, especially for low-level abnormalities.
- Further optimization may be needed for flagging specific abnormal cell types and band neutrophils.