Accurate analysis of fluorescence decays from single molecules in photon counting experiments
David A Turton1, Gavin D Reid, Godfrey S Beddard
1Department of Chemistry, University of Leeds, Leeds LS2 9JT, UK.
Abstract:
In time-resolved, single-photon counting experiments, the standard method of nonlinear least-squares curve fitting incorrectly estimates the fluorescence lifetimes. Even for single-exponential data, errors may be up to +/- 15%, and for more complex fits, may be even higher, although the fitted line appears to describe the data. The origin of this error is not a result of the Poisson distribution, as is often stated, but is entirely due to the weighting of the fit. An alternative weighting method involving a minor change in the fitting method eliminates this problem, enabling accurate fitting even in difficult cases, including the small data sets observed in single molecule experiments and with a precision similar to that of maximum likelihood methods.
Related Concept Videos
Protein Dynamics in Living Cells
Fluorescent recovery after photobleaching (FRAP) is a fluorescent-protein-based detection technique used to quantify protein movement rates within the cell. This method exposes a small portion of the cell to an intense laser beam. The laser beam causes permanent photobleaching of the fluorophore-tagged proteins in the exposed region. As the bleached...