Related Experiment Video
Updated: Aug 30, 2026

Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models
Published on: November 11, 2020
[Aerosol deposition and clinical performance verified with a spacer device made in Brazil]
P A Camargos1, J A Rubim, C J Simal
1Universidade Federal de Minas Gerais, Belo Horizonte, MG, Brazil.
Abstract:
OBJECTIVE: To assess the lung deposition pattern of radioaerosol and the clinical performance of a spacer developed and made in Brazil. METHODS: Qualitative - in a patient with cystic fibrosis - and semi-quantitative - in two healthy volunteers - assessment of pulmonary deposition of (99)mtechnetium was done using the Aerogama Medical oxigen driven nebulizer system attached to the spacer and a gama-camera (Siemens, model Orbiter) connected to a microcomputer. In the next step, clinical assessment was carried out in 50 asthmatic children, aged from four months to 13 years old with an acute attack, using conventional doses of albuterol through a metered dose inhaler attached to the spacer device. RESULTS: Qualitative assessment revealed a lung silhouette comparable with those obtained in the inhalation scintigraphy and semiquantitative assessment reveals that 7.5% to 8.0% of the inhaled (99m)technetium reached the volunteerś lungs. Statistically significant differences (p < 0.001) were observed comparing clinical scores at admission with those verified 20 and 40 minutes after albuterol inhalation; conversely, no significance was obtained for scores taken at 60 and 80 minutes. CONCLUSIONS: Although we used an alternative method, the scintigraphic assessment reveals an expected pattern of pulmonary deposition. Similarly, clinical performance in the treatment of an acute attack showed results comparable with those obtained with other spacers devices.

