Study Objective:
To compare the frequency of airway seal and sore throat with the LMA-ProSeal (PLMA) and the standard Laryngeal Mask Airway (LMA) during laparoscopic surgery.
Design:
Prospective, controlled, randomized, nonblinded clinical study.
Setting:
University-affiliated hospital.
Patients:
60 adult, ASA physical status I, II, and III patients undergoing laparoscopic surgery with general anesthesia, without contraindication to the use of the laryngeal mask.
Interventions:
Patients were randomized to receive mechanical ventilation [tidal volume (V(T)) 7 mL/kg(-1); positive end-expiratory pressure (PEEP) 10 cmH(2)O] through the PLMA or the standard LMA, both equipped with a gastric tube.
Measurements:
Heart rate, arterial pressure, inspiratory and expiratory V(T), airway pressure, end-tidal CO(2) partial pressure, and pulse oximetry were recorded. The leak fraction was calculated as the difference between the inspiratory and expiratory V(T) divided by the inspiratory V(T). Postoperative sore throat frequency was scored in the recovery room ("early") and 1 week after surgery ("late").
Main Results:
All patients were successfully ventilated through the assigned laryngeal mask. The leak fraction was 7 +/- 3% with the LMA and 7 +/- 4% with the PLMA (p = 0.731). In one patient, the PLMA drainage tube was not patent despite a leak fraction of 5%, and there was no clinically detectable air leak. During the recovery room stay, the frequency of sore throat was scored as mild in 13% and 10% of patients with the standard LMA and the PLMA, respectively, and was absent in the remaining patients (p = 0.99, between groups). There were no differences in the frequency of sore throat between the "early" and "late" evaluations (p = 0.99).
Conclusions:
The PLMA and the LMA show similar airtight efficiency during laparoscopy. The patency of the PLMA drainage tube should always be confirmed. The sore throat evaluation performed in recovery room appears as reliable as later evaluations.
Related Concept Videos
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques