Evaluation of imprecision for cardiac troponin assays at low-range concentrations
Mauro Panteghini1, Franca Pagani, Kiang-Teck J Yeo
1Laboratorio Analisi Chimico Cliniche 1, Azienda Ospedaliera 'Spedali Civili', 25125 Brescia, Italy. pantegi@bshosp.osp.unibs.it
Insights
No cardiac troponin assay met the recommended imprecision (CV) of 10% at the 99th percentile decision limit for myocardial infarction (MI). This study evaluated commercially available assays, finding significant variability in performance. Further research is needed to improve assay accuracy.
Area of Science:
- Clinical Chemistry
- Cardiovascular Diagnostics
- Biomarker Assay Validation
Background:
- The 99th percentile of cardiac troponin is the recommended threshold for diagnosing myocardial infarction (MI).
- A coefficient of variation (CV) of ≤10% at this decision limit is advised for assay imprecision.
- Limited peer-reviewed data exists on the imprecision of commercially available cardiac troponin assays.
Purpose of the Study:
- To evaluate the imprecision profiles of various commercial cardiac troponin assays.
- To determine the lowest concentration with a 10% CV for each assay.
- To assess assay performance against the recommended MI diagnostic criteria.
Main Methods:
- Eight serum pools with varying cardiac troponin concentrations were prepared and stored.
- Measurements were performed using a standardized protocol: two replicates per specimen, one run daily for 20 days.
- Assays were run using two reagent lots and three calibrations, with data analyzed per NCCLS EP5-A guidelines.
Main Results:
- Significant variability in the lowest concentration achieving a 10% CV was observed across assays.
- Notable results included Elecsys (0.04 µg/L), Access (0.06 µg/L), and Liaison (0.065 µg/L).
- Other assays like AxSYM (1.22 µg/L) and Opus (0.90 µg/L) showed higher concentrations for 10% CV.
Conclusions:
- None of the evaluated cardiac troponin assays met the recommended 10% CV at the manufacturer-defined 99th percentile reference limit.
- This highlights a critical gap in assay performance for accurate MI diagnosis.
- Assay imprecision remains a significant challenge in adhering to current MI diagnostic guidelines.
Background:
The European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction (MI) has recommended that an increased cardiac troponin should be defined as a measurement above the 99th percentile value of the reference group. A total imprecision (CV) at the decision limit of =10% is recommended. However, peer-reviewed data on assay imprecision are lacking. The purpose of this study was to construct the clinically relevant imprecision profiles for each of the commercially available cardiac troponin assays. Pools of human sera containing cardiac troponin concentrations around the MI decision limit were assessed to identify the lowest concentration associated with a 10% CV.
Methods:
Eight serum pools targeting different concentrations of cardiac troponin (I and T) were prepared and stored at -70 degrees C until usage. The cardiac troponin measurement protocol consisted of two replicates per specimen per run, and one run per day for 20 days, using two reagent lots and three calibrations. Manufacturers of each cardiac troponin assay directly performed the measurements. Data analysis for each assay was centralized and performed according to the NCCLS EP5-A guideline.
Results:
The lowest concentrations (microg/L) providing a 10% CV were as follows: AxSYM, 1.22; ACS:180, 0.37; Centaur, 0.33; Immuno 1, 0.34; Access, 0.06; Vidas, 0.36; Liaison, 0.065; Dimension, 0.26; Opus, 0.90; Stratus CS, 0.10; Immulite, 0.32; Vitros ECi, 0.44; Elecsys, 0.04; AIA 21, 0.09.
Conclusion:
No cardiac troponin assay was able to achieve the 10% CV recommendation at the 99th percentile reference limit defined by the manufacturer.
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Myocarditis II: Clinical Features and Diagnostic Tests
Acute Coronary Syndrome III: Diagnostic Studies