Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
[Systematics of imaging artifacts in MRT caused by metallic vascular implants (stents)]
1Sektion für Experimentelle Radiologie, Eberhard-Karls-Universität Tübingen, Tübingen. hansjoerg.graf@med.uni-tuebingen.de
Purpose:
Imaging artifacts in magnetic resonance tomography (MRT) caused by metallic vascular implants (stents) were characterized systematically in dependence on the material and the construction of the implants as well as with respect to different measurement protocols and for different static field strength B (0).
Methods:
Twelve stents, different in material (stainless steel, nitinol, Co-alloy) and/or construction, were examined at B (0) = 0.2 T and 1.0 T with two 2D gradient echo (GE) sequences (TE = 4 and 10 ms), one 3D GE sequence and one spin echo (SE) sequence. The stents were put into water with Gd-DTPA contrast agent. The dependence on the orientation was analyzed for the 9 possibilities of an orthogonal alignment of the stent axis, the direction of B (0), the slice, the read out, and the phase encoding direction. Special interest was given to the visibility of the stent lumen at forced rf excitation, as well as to the influence of broken struts on the signal obtained.
Results:
For the examined stents GE technique showed, due to spin dephasing regions a total signal loss ranging up to 8 mm away from the stent mashes. The value depends on the stent material, the stent orientation in the scanner and grows with voxel size, echo time and B (0). In SE technique dephasing artifacts vanish and wrong spatial encoding gets visible, rf shielding is more pronounced. The visibility of the stent lumen ranges from nearly unperturbed down to a complete signal loss. An improvement is possible using enlarged flip angles. Broken struts can not be imaged significantly.
Discussion:
The MR representation of metallic stents commercially available at the time, especially of nitionol stents, can be optimized with a suitable adaptation of the imaging parameters. However, a profound improvement can only be expected from new stent material and design.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies IV: Magnetic Resonance Imaging