Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
A 50% n-octylmethyl, 50% diphenyl-polysiloxane as stationary phase with unique selectivity for gas chromatography
Bernhard X Mayer1, Hanspeter Kählig, Walfried Rauter
1Department of Clinical Pharmacology, University School of Medicine, Vienna University, Währinger Gürtel 18-20, 1090 Vienna, Austria. Bernhard.Mayer@univie.ac.at
Abstract:
An n-octylmethyl, diphenyl-polysiloxane called SOP-50-Octyl was prepared by a condensation reaction of bis(dimethylamino) n-octylmethylsilane with diphenylsilanediol. The resulting copolymer was a gum with high molecular weight and was obtained with a yield of 80%. 1H and 29Si NMR spectroscopy revealed that the copolymer was a 52% octylmethyl, 48% diphenyl-polysiloxane with random microstructure. Small cyclic impurities could be almost quantitatively removed via a purification step. SOP-50-Octyl was used as stationary phase for the preparation of wall coated open tubular fused silica capillary columns for gas-liquid chromatography. The capillary columns exhibited high separation efficiency and high inertness. The stationary phase offered a unique selectivity due to its unique composition. The Rohrschneider-McReynolds constants indicated a low overall polarity in spite of the high phenyl content, as the polarity was distinctly decreased by the octyl substituent. Furthermore, the octyl substituent was responsible for a high column bleed, reducing the maximum allowable operating temperature to 280 degrees C. The elution temperatures of apolar compounds were increased due to increased interaction of the octyl substituent with the analytes. Some applications with volatile and semi-volatile organic compounds illustrate that SOP-50-Octyl is an excellent choice for confirmational analyses.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Size-Exclusion Chromatography