Detection of human polyomaviruses in urine from bone marrow transplant patients: comparison of electron microscopy
Stefan S Biel1, Andreas Nitsche, Andreas Kurth
1Beiersdorf AG, Hamburg, Germany. stefan.biel@beiersdorf.com
Background:
We studied electron microscopy (EM) as an appropriate test system for the detection of polyomavirus in urine samples from bone marrow transplant patients.
Methods:
We evaluated direct EM, ultracentrifugation (UC) before EM, and solid-phase immuno-EM (SPIEM). The diagnostic accuracy of EM was studied by comparison with a real-time PCR assay on 531 clinical samples.
Results:
The detection rate of EM was increased by UC and SPIEM. On 531 clinical urine samples, the diagnostic sensitivity of EM was 47% (70 of 149) with a specificity of 100%. We observed a linear relationship between viral genome concentration and the proportion of urine samples positive by EM, with a 50% probability for a positive EM result for urine samples with a polyomavirus concentration of 10(6) genome-equivalents (GE)/mL; the probability of a positive EM result was 0% for urine samples with <10(3) GE/mL and 100% for urine samples containing 10(9) GE/mL.
Conclusions:
UC/EM is rapid and highly specific for polyomavirus in urine. Unlike real-time PCR, EM has low sensitivity and cannot quantify the viral load.