A new rigid biodegradable anchor for meniscus refixation: biomechanical evaluation
Thore Zantop1, Anne Kathleen Eggers, Volker Musahl
1Department of Orthopaedic Surgery, Christian-Albrechts-University Kiel, Michaelisstrasse 1, 24105 Kiel, Germany.
Summary
The Contour Meniscus Arrow offers improved biomechanical properties compared to the original Meniscus Arrow implant for meniscus repair. This second-generation device shows comparable strength to traditional sutures, justifying its clinical use.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Sports medicine
Background:
- Conventional meniscus repair sutures can cause tissue damage and prolong operating times.
- All-inside repair devices aim to mitigate these disadvantages.
- The Contour Meniscus Arrow is an advanced biodegradable implant for meniscus repair.
Purpose of the Study:
- To compare the biomechanical properties of the Contour Meniscus Arrow with its predecessor, the Meniscus Arrow.
- To evaluate the Contour Meniscus Arrow against traditional vertical and horizontal meniscus suture techniques.
- To assess fixation strength, stiffness, and failure modes of different meniscus refixation methods.
Main Methods:
- Utilized fresh frozen bovine menisci for testing.
- Employed a computer-based material-testing machine at a rate of 12.5 mm/s.
- Evaluated four refixation techniques: Meniscus Arrow, Contour Meniscus Arrow, vertical sutures, and horizontal sutures.
Main Results:
- Vertical sutures exhibited the highest initial fixation strength, followed by horizontal sutures and the Contour Arrow.
- The original Meniscus Arrow demonstrated inferior pull-out strength.
- Cyclic testing reduced fixation strength across all tested groups.
- The Contour Arrow showed superior pull-out strength and comparable stiffness to the original Meniscus Arrow.
Conclusions:
- The Contour Meniscus Arrow demonstrates improved or comparable biomechanical performance to previous all-inside devices.
- Its pull-out strength is similar to reported values for horizontal meniscus sutures.
- The biomechanical profile of the Contour Meniscus Arrow supports its consideration for clinical application in meniscus repair.