Test-retest reliability of clinical Shack-Hartmann measurements
Xu Cheng1, Nikole L Himebaugh, Pete S Kollbaum
1School of Optometry, Indiana University, Bloomington, Indiana 47405, USA. xcheng@indiana.edu
Purpose:
To evaluate the stability of clinical monochromatic aberrometry measurements over a wide range of time scales.
Methods:
Monochromatic aberrations in four normal eyes were measured with a clinical Shack-Hartmann aberrometer. A chin rest or a supplemental bite bar attachment was used to stabilize head and eye position. Five repeated measurements were taken within one test (5 frames, t < 1 second) without realignment. With realignment between each measurement, aberration measurements were repeated five times (t < 1 hour) on each day, at the same time of day on five consecutive days, and again on 5 days at monthly intervals. A control experiment studied the effect of systematically misaligning the eye to determine whether fixation errors can account for the variation in the repeated measurements.
Results:
Variability of wavefront root mean square (RMS) error (excluding defocus and astigmatism) was tracked across repeated measurements. Variances for different time scales were: 8.10 x 10(-5) microm2 (t < 1 second), 3.24 x 10(-4) microm2 (t < 1 hour), 4.41 x 10(-4) microm2 (t < 1 week), 9.73 x 10(-4) microm2 (t < 1 year). Bite bar and chin rest data were almost identical. Rotational fixation error up to 3 degrees accounts for only part of the variability.
Conclusions:
Increased variability in aberration maps between days and months indicates biological fluctuations that are large enough to prevent achievement of "perfect vision," even in the unlikely event that spherical and astigmatic refractive errors are corrected perfectly. However, lack of stability does not justify withholding treatment. A lasting benefit of aberration correction is expected despite temporal variability.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:18Doppler Ultrasound-Based Leg Blood Flow Assessment During Single-Leg Knee-Extensor Exercise in an Uncontrolled Setting
Published on: December 15, 2023
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Comparing Experimental Results: Student's t-Test