Luís Eduardo Silva Soares1, Aírton Abrahão Martin, Antonio Luiz Barbosa Pinheiro
1School of Dentistry, Research and Development Institute, IP&D, UNIVAP, São José dos Campos, SP, Brazil. lesoares@univap.br
Objective:
Near infrared Raman spectroscopy (RS) was used to monitor, in vitro, the degree of conversion (DC) of composite resins (Z100, 3M), photoactivated by both the halogen lamp and the argon laser beam.
Background Data:
Several methods were used to study the alterations of composite resins. Vibration methods such as RS allow a precise assessment of the depth of polymerization and the degree of conversion of composite resins.
Materials And Methods:
Sixty circular blocks of resin (7 mm x 2.5 mm) were cured using a halogen light source (n=30, lambda=400-500 nm, power density=478 mW/cm2) or an argon laser beam (n=30, lambda=488 nm, power density=625 mW/cm2) using the same irradiation time (5, 10, 20, 30, 40, and 60 sec). The directly irradiated (top) and the non-irradiated (bottom) surfaces were analyzed immediately after curing by Raman spectroscopy.
Results:
The Raman results show systematic changes of the relative intensities between the peaks at 1610 (aromatic C=C stretching mode) and the 1640 cm(-1) (methacrylate C=C stretching mode), as a function of irradiation time. After 60 sec of irradiation time, the maximum degree of conversion reached for the samples cured either by the argon laser or halogen lamp was 66.4% and 62.2%, respectively.
Conclusion:
The argon laser was more effective and showed better biocompatibility, with less residual monomer in the bottom (2.5 mm). These results show that RS can be used as an effective method to study the degree of conversion of composite resins.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...