Background:
The Symmetry aortic connector uses a nitinol implant to create proximal anastomoses with saphenous vein grafts. Multiple detector-row cardiac computed tomography (MDCT) is used as a noninvasive method of quality control at our institution.
Methods:
In 50 elective coronary artery bypass grafting patients who participated in a randomized trial comparing automated with conventionally hand-sewn proximal anastomoses, MDC T was performed on postoperative day 5. Fifty-three automated Symmetry anastomoses were created in 34 patients (group 1). Twenty-five conventionally hand-sewn anastomoses created in 16 patients served as controls (group 2). Graft patency and the presence or absence of high-grade stenosis at the proximal anastomotic site were evaluated.
Results:
In group 1, 2 (3.8%) of the grafts were found occluded at MDCT or coronary angiography, and no further relevant stenosis was observed. In group 2, at postoperative MDCT all grafts were found patent without significant narrowing of the proximal anastomotic site.
Conclusions:
The feasibility of proximal anastomoses using the Symmetry device has been reported. Patency control with invasive angiography has been performed by other groups. With MDCT, noninvasive evaluation of proximal anastomotic quality and graft patency is possible, even if nitinol is implanted.
More Related Videos
12:00Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
04:35Sample Preparation for Computed Tomography-based Three-dimensional Visualization of Murine Hind-limb Vessels
Published on: October 7, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT