Influence of implant length and diameter on stress distribution: a finite element analysis
Lucie Himmlová1, Tat'jana Dostálová, Alois Kácovský
1Institute of Dental Research and Czech Technical University, Prague, Czech Republic. himmlova@seznam.cz
Statement Of Problem:
Masticatory forces acting on dental implants can result in undesirable stress in adjacent bone, which in turn can cause bone defects and the eventual failure of implants.
Purpose:
A mathematical simulation of stress distribution around implants was used to determine which length and diameter of implants would be best to dissipate stress.
Material And Methods:
Computations of stress arising in the implant bed were made with finite element analysis, using 3-dimensional computer models. The models simulated implants placed in vertical positions in the molar region of the mandible. A model simulating an implant with a diameter of 3.6 mm and lengths of 8 mm, 10 mm, 12 mm, 14 mm, 16 mm, 17 mm, and 18 mm was developed to investigate the influence of the length factor. The influence of different diameters was modeled using implants with a length of 12 mm and diameters of 2.9 mm, 3.6 mm, 4.2 mm, 5.0 mm, 5.5 mm, 6.0 mm, and 6.5 mm. The masticatory load was simulated using an average masticatory force in a natural direction, oblique to the occlusal plane. Values of von Mises equivalent stress at the implant-bone interface were computed using the finite element analysis for all variations. Values for the 3 most stressed elements of each variation were averaged and expressed in percent of values computed for reference (100%), which was the stress magnitude for the implant with a length of 12 mm and diameter of 3.6 mm.
Results:
Maximum stress areas were located around the implant neck. The decrease in stress was the greatest (31.5%) for implants with a diameter ranging from of 3.6 mm to 4.2 mm. Further stress reduction for the 5.0-mm implant was only 16.4%. An increase in the implant length also led to a decrease in the maximum von Mises equivalent stress values; the influence of implant length, however, was not as pronounced as that of implant diameter.
Conclusions:
Within the limitations of this study, an increase in the implant diameter decreased the maximum von Mises equivalent stress around the implant neck more than an increase in the implant length, as a result of a more favorable distribution of the simulated masticatory forces applied in this study.
Related Concept Videos
Stress Concentrations in Circular Shafts
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to its distance...
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes.
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Saint-Venant's Principle
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity