Related Experiment Videos
Comparative study of capillary zone electrophoresis and high-performance liquid chromatography in the analysis of
P J Oefner1, G K Bonn, C G Huber
1Department of Analytical Chemistry, Johannes-Kepler-University, Linz, Austria.
Abstract:
Capillary zone electrophoresis and high-performance liquid chromatography were compared with regard to the separation of oligonucleotides and double-stranded DNA. Both anion-exchange and reversed-phase high-performance liquid chromatography on non-porous particles are considered to be superior to capillary electrophoresis in terms of speed and selectivity in the analysis of oligonucleotides up to 30 bases in length. Moreover, reversed-phase chromatography allows the simultaneous purification of detritylated oligonucleotides with recoveries > 90%. Compared with anion-exchange chromatography, there is no need for a subsequent desaltation step because the volatile buffer system can be readily evaporated. With regard to dsDNA, however, the resolving power of capillary electrophoresis cannot be matched by anion-exchange chromatography at present. Moreover, the combined use of hydroxyethylcellulose and ethidium bromide not only yielded a separation efficiency equal to that achieved by means of gel-filled capillaries but also avoids some of their limitations such as the destruction of the gel matrix at high current densities and the bias involved in electrokinetic injection.