Assessment of motion gating strategies for mouse magnetic resonance at high magnetic fields

Paul J Cassidy1, Jürgen E Schneider, Stuart M Grieve

  • 1Oxford Cardiac Research Group Magnetic Resonance Unit, University Laboratory of Physiology, University of Oxford, Oxford, UK. paul.cassidy@physiol.ox.ac.uk

Abstract

Insights

User-defined physiological gating significantly improves mouse cardiac magnetic resonance (MR) imaging and spectroscopy at high magnetic fields. This method enhances signal intensity and reduces noise for better in vivo results.

Area of Science:

  • Biomedical Engineering
  • Magnetic Resonance Imaging
  • Cardiovascular Research

Background:

  • High magnetic fields in mouse cardiac MRI offer enhanced resolution but are susceptible to motion artifacts.
  • Effective motion gating is crucial for acquiring high-quality in vivo cardiac images and spectra in small animal models.

Purpose of the Study:

  • To evaluate the efficacy of various motion gating strategies for mouse cardiac magnetic resonance (MR) at 11.75 T.
  • To quantify motion artifact levels in cardiac MR images and spectra using different gating techniques.

Main Methods:

  • Performed MR imaging, angiography, and 1H-spectroscopy on anesthetized C57Bl/6 mice at 11.75 T.
  • Utilized a custom physiological motion gating device to compare no gating, cardiac gating, conventional gating, automatic gating, and user-defined gating.
  • Assessed gating performance by quantifying image motion artifacts and spectral amplitude/phase stability.

Main Results:

  • User-defined gating with steady-state maintenance during respiration demonstrated superior performance.
  • Achieved a threefold increase in signal intensity and a sixfold reduction in noise intensity compared to cardiac gating alone.
  • User-defined gating optimized cardiac imaging, angiography, and spectroscopy.

Conclusions:

  • Physiological gating incorporating steady-state maintenance during respiration is essential for high-field mouse cardiac MR.
  • This optimized gating strategy significantly improves image and spectral quality in vivo.
  • The findings are critical for advancing small animal cardiovascular research using high-field MRI.