Pacemaker malfunction due to subcutaneous emphysema--a case report.

M Santomauro1, S Ferraro, G Maddalena

  • 1Department of Cardiology and Cardiac Surgery, 2nd Medical School, University of Naples, Italy.

Angiology
|October 1, 1992
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Distinct molecular and clinical aggressiveness in very early-onset metastatic colorectal cancer: survival and genomic divergence between patients aged 30-39 versus 40-49 years.

ESMO open·2026
Same author

Immune checkpoint inhibitors for POLE or POLD1 proofreading-deficient metastatic colorectal cancer.

Annals of oncology : official journal of the European Society for Medical Oncology·2024
Same author

Defining the subset of mutations in polymerase epsilon (POLE) associated with loss-of-proofreading (LOP) functionality.

Annals of oncology : official journal of the European Society for Medical Oncology·2024
Same author

An update on atrial fibrillation in 2014: From pathophysiology to treatment.

International journal of cardiology·2015
Same author

Intra-aortic balloon pump: towards continuous monitoring of control parameters.

Minerva anestesiologica·2013
Same author

Resynthesis of sternal dehiscence with autologous bone graft and autologous platelet gel.

Journal of wound care·2012

Pacemaker malfunction occurred due to increased impedance caused by air in the pacemaker pocket. This case highlights a rare complication of pacemaker implantation.

Area of Science:

  • Cardiology
  • Medical Devices
  • Implantable Electronic Devices

Background:

  • Pacemakers are crucial for managing cardiac arrhythmias.
  • Pacemaker pocket complications can affect device function.
  • Impedance monitoring is vital for pacemaker integrity.

Observation:

  • A case of pacemaker malfunction was reported.
  • The malfunction was linked to a critical rise in impedance.
  • Air entrapment within the pacemaker pocket was identified as the cause.

Findings:

  • Increased pocket impedance due to air entrapment can lead to pacemaker failure.
  • This specific mechanism of malfunction is uncommon.
  • Diagnostic evaluation identified the air pocket as the source of impedance increase.

Related Experiment Videos

Implications:

  • Awareness of air entrapment as a cause of pacemaker malfunction is important for clinicians.
  • Proper surgical technique may help prevent air pocket formation.
  • Further investigation into impedance monitoring for early detection of such complications is warranted.