Measurement of bone conduction levels for high frequencies
Martin L Lenhardt1, Douglas G Richards, Alan G Madsen
1Sound Technique Systems, LLC, Program in Biomedical Engineering, Virginia Commonwealth University, Richmond, VA 23298-0168, USA. lenhardt@hsc.vcu.edu
Abstract:
For assessment of safety, it is necessary to measure the maximum possible force exerted by a bone conduction device coupled to the human head. Calibration of bone conduction hearing aids and vibrators in the audiometric range is based on measurement of acceleration and force using an artificial mastoid. Extending the measurement to the high audio range was accomplished using a live head. To assess safety of the UltraQuiet tinnitus treatment system, as an example, acceleration was measured from 5 to 20 kHz on a live human head as compared with calibrated levels at 6 kHz on an artificial mastoid and the live head. Using head acceleration and anchoring it to established calibration levels is a means of establishing clinical safety. Stimulation in the high audio frequencies at low levels was found to be safe. In contrast, stimulation with ultrasound requires more energy (approximately 75-90 dB re 6 kHz), which may increase the risk of damage to the car.
Related Concept Videos
The Cochlea
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and hence a...
Perceiving Loudness, Pitch, and Location
Place theory, or place coding, suggests that different pitches are heard because various sound waves activate specific locations along the cochlea's basilar membrane. The brain determines the pitch of a sound by identifying...