Laboratory and field validations of a solid-phase microextraction device for the determination of ethylene oxide
Shih-Wei Tsai1, Shih-Ting Tsai, Ven-Shing Wang
1Department of Occupational Safety and Health, Institute of Environmental Medicine, China Medical University, 91 Hsueh-Shih Road, Taichung, Taiwan.
Abstract:
Laboratory and field evaluations were performed to validate a solid-phase microextraction (SPME) device that was used as a diffusive sampler. Hydrogen bromide (HBr) was loaded onto the carboxen-polydimethylsiloxane (CAR-PDMS) fiber for the determination of ethylene oxide (EtO) with on-fiber derivatization. For laboratory evaluations, known concentrations of ethylene oxide around the threshold limit values (TLV)/time-weighted average and specific relative humidities (RHs) were generated by syringe pumps in a dynamic generation system. The SPME diffusive samplers and the commercially available 3M 3551 passive monitors were placed side-by-side in an exposure chamber which was designed to allow measurement of face velocities, temperatures, exposing vapor concentrations, and RHs. Field validations with both SPME diffusive sampler and 3M 3551 passive monitors were also performed. The correlations between the results from both SPME device and 3M 3551 passive monitor were found to be linear (r > 0.9699) and consistent (slope approximately equal to 1.12 +/- 0.07). However, the variations of diffusion coefficients at different temperatures needs to be considered and the adjustment of sampling constant was a must when sampling at temperatures different from 25 degrees C.
More Related Videos
07:26A Simple Method for Automated Solid Phase Extraction of Water Samples for Immunological Analysis of Small Pollutants
Published on: January 1, 2016
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024