Criterion validity, test-retest reliability and sensitivity to change of the St George Urinary Incontinence Score
A L Blackwell1, W Yoong, K H Moore
1Pelvic Floor Unit, St George Hospital, University of New South Wales, Sydney, Australia.
Objective:
To assess the criterion validity, test-retest reliability and the sensitivity to change after treatment of the St George Urinary Incontinence Score (SGUIS).
Patients And Methods:
Women presenting with urinary incontinence completed two SGUIS tests one week apart. A frequency-volume chart (FVC) was completed before the first attendance and the women had a 1-h pad test before treatment. Patients were treated conservatively under the care of a urogynaecologist or nurse continence advisor, or surgically with an open or laparoscopic colposuspension. After treatment, the SGUIS, FVC and 1-h pad test were repeated.
Results:
The SGUIS correlated moderately well with the number of leaks per week (Spearman's r = 0.610, 95% confidence interval 0.516-0.689, P < 0.001) but less well with the 1-h pad test loss (r = 0.257, 0.124-0.380, P = 0.002). The test-retest reliability was acceptable, as the mean (SD) difference between the first and second SGUIS was 0.337 (2.675), with limits of agreement of -5.012 to -5.686. The change in the SGUIS after treatment correlated well with the improvement in the number of leaks per week (r = 0.742, 0.662-0.805, 156 samples, P < 0.001) but not as well with the change in 1-h pad test loss (r = 0.531, 0.405-0.636, 151, P < 0.001), although the trend was similar.
Conclusion:
The criterion validity of the SGUIS appeared adequate for both the number of leaks per week on the FVC and leakage on the 1-h pad test. The statistical reproducibility of the test was adequate and appeared more responsive to change after treatment than the other measures. The score is suitable for rapid self-administration by patients with a range of incontinence types, unlike most other currently available test instruments.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:18Doppler Ultrasound-Based Leg Blood Flow Assessment During Single-Leg Knee-Extensor Exercise in an Uncontrolled Setting
Published on: December 15, 2023
Related Concept Videos
Reliability and Validity
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Urodynamic Studies: Uroflowmetry
Urine Studies II: Urine Culture and Sensitivity Test