Wedge filter effects on dosimetric parameters of a linear accelerator
Summary
Output factors for wedged and open fields were measured for two linear accelerators. One accelerator showed significant output variations (up to 5%) with wedge filters, unlike the other which had minimal changes.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Wedge filters are crucial in radiation therapy for shaping dose distributions.
- Understanding their impact on dosimetric parameters is essential for accurate treatment planning.
Purpose of the Study:
- To measure and compare open-field and wedged-field output factors for two different linear accelerators.
- To investigate the variability of wedge factors with field size and their effect on other dosimetric parameters.
Main Methods:
- Measured output factors as a function of field size for open and wedged fields on two linear accelerators.
- Determined wedge factors by calculating the ratio of output with and without the wedge filter.
- Analyzed variations in output factors and wedge factors across different field sizes and wedge angles.
Main Results:
- One linear accelerator exhibited output factor differences up to 5% between wedged and open fields, with wedge factors varying by field size.
- The second linear accelerator showed less than 1% variation in output factors and wedge factors (for a 60-degree wedge) across field sizes.
- Wedge filters were confirmed to alter isodose distributions, percent depth dose curves, output factors, and wedge factors.
Conclusions:
- The impact of wedge filters on dosimetric parameters varies significantly between different linear accelerators.
- Accurate characterization of wedge effects is critical for ensuring precise radiation delivery and patient safety.