Integrated software for the analysis of brain PET/SPECT studies with partial-volume-effect correction.

Mario Quarantelli1, Karim Berkouk, Anna Prinster

  • 1Biostructure and Bioimaging Institute, National Council for Research, Building 10, Via Pansini 5, 80131 Naples, Italy. quarante@unina.it

Journal of Nuclear Medicine : Official Publication, Society of Nuclear Medicine
|February 13, 2004
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Higher amyloid deposition and lower white matter volume in cognitively healthy seniors exceeding French alcohol recommendations.

EBioMedicine·2026
Same author

Amyloid PET Quantitation and Centiloid Thresholds in the Diagnosis of Alzheimer Disease: An Individual Participant Data Meta-Analysis.

JAMA·2026
Same author

Structural and molecular determinants of medial temporal lobe network vulnerability in aging and Alzheimer's disease.

Alzheimer's research & therapy·2026
Same author

An automated quantitative report for multiple sclerosis using only 3D T2-fluid-attenuated inversion recovery MRI.

Neuroradiology·2026
Same author

Ultra-high-field fMRI study of insular discrimination of taste quality and perceived taste valence.

Chemical senses·2026
Same author

Night-to-night rapid eye movement sleep variability: A relevant marker of early amyloid-β deposition.

Alzheimer's & dementia : the journal of the Alzheimer's Association·2026

This study introduces software for integrated brain PET/MRI analysis, improving partial-volume-effect correction (PVEc) for accurate gray matter quantification. The R-PVEc and mMG-PVEc methods significantly enhance accuracy, especially in the presence of atrophy and coregistration errors.

Area of Science:

  • Neuroimaging
  • Medical Physics
  • Biomedical Engineering

Background:

  • Partial-volume effects (PVEc) in brain PET studies lead to inaccurate quantification, particularly in the presence of atrophy.
  • Existing methods for PVEc often have limitations in accuracy and precision.
  • Integrated analysis of PET and MRI data is crucial for robust brain imaging studies.

Purpose of the Study:

  • To present novel software for integrated analysis of brain PET and MRI data.
  • To evaluate four different partial-volume-effect correction (PVEc) methods using simulated PET studies.
  • To assess the accuracy and precision of PVEc methods under varying degrees of atrophy and experimental errors.

Main Methods:

  • Software developed for automated region of interest (ROI) placement on coregistered segmented MRI.

Related Experiment Videos

  • Application of four distinct PVEc techniques to simulated (18)F-FDG PET studies.
  • Introduction of experimental errors including coregistration, segmentation, and resolution estimates to simulate realistic conditions.
  • Main Results:

    • Uncorrected PET values showed significant underestimation of gray matter (GM) and overestimation of white matter (WM).
    • Voxel-based correction improved GM values but residual underestimation persisted.
    • R-PVEc and mMG-PVEc methods achieved over 96% accuracy, with mMG-PVEc offering the lowest coefficient of variation (6.0%) for GM ROIs.
    • Coregistration errors were identified as the primary source of imprecision.

    Conclusions:

    • Integrated automated ROI placement and PVEc software enables accurate recovery of true GM ROI values in brain PET/MRI.
    • R-PVEc and mMG-PVEc methods demonstrate high accuracy, with mMG-PVEc being suitable for corrected image generation.
    • The developed software provides a valuable tool for precise quantitative analysis in neuroimaging research.