Temperature dependence of spurious pulses in use of plastic scintillation detectors
Yasushi Kawada1, Jun Ito, Qiu-Wei Wang
1Japan Radioisotope Association, 28-45 Honkomagome 2-cyome, Bunkyo-ku, Tokyo 113-8941, Japan. kawada@jrias.or.jp
Abstract:
Pulse-height and time spectra of afterpulses from a scintillation detector with NE102A scintillator sandwiching a 60Co source were observed at several different temperatures. From time analysis with a slow amplifier followed by a conventional TAC, spurious pulses are grouped into two types; time-dependent afterpulses and random noise. The intensity of afterpulses decreased considerably with increasing temperature, while the intensity of random noise pulses increased abruptly above 50 degrees C. Experiments, in which the temperatures of scintillator sheets and of the photomultiplier tube were changed separately, suggested that the time-dependent afterpulses were produced in the scintillator itself, while the random component arose from photomultiplier tube.