Purpose:
To develop an automatic registration method for electrocardiogram-gated myocardial perfusion single-photon emission computed tomography (SPECT) and cardiac cine-magnetic resonance imaging (MRI).
Materials And Methods:
Paired myocardial perfusion SPECT (MPS) and MRI from 20 patients were considered. MR images were presegmented by heart localization based on detection of cardiac motion and optimal thresholding. A registration algorithm based on mutual information was subsequently applied to all time frames or a selected subset from both modalities.
Results:
A preprocessing step significantly improved the accuracy of the registration when compared to automatic registration performed without preprocessing. Errors in translation parameters (T(x), T(y), T(z)) averaged (1.0 +/- 1.5, 1.1 +/- 1.3, 0.9 +/- 0.9) pixels with MRI segmentation and (4.6 +/- 3.2, 3.4 +/- 2.6, 3.0 +/- 3.4) pixels without MRI segmentation. Errors in rotation parameters (R(x), R(y), R(z)) averaged (5.4 +/- 2.9, 3.4 +/- 2.7, 4.5 +/- 3.6) degrees with MRI segmentation and (9.3 +/- 6.1, 4.8 +/- 4.3, 14.6 +/- 12.6) degrees without MRI segmentation. Error was calculated as the absolute difference between the expert manual and the automatic registration transformation.
Conclusion:
Automatic registration of gated MPS and cine MRI is possible with the use of a mutual information-based technique when MR images are presegmented. Cardiac motion can be used to isolate the left ventricle (LV) on the MR images automatically, and subsequently the segmented MR images can be coregistered with gated MPS.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
06:57Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI