Test-retest reliability of the commercial volumeter
Rachel L Dodds1, Kelly A Nielsen, Andrea G Shirley
1School of Occupational Therapy, University of Indianapolis, Indianapolis, IN 46227, USA.
Objective:
The purpose of this study was to improve the standardization of the current protocol for measuring hand volume with the commercial volumeter.
Method:
A repeated-measures design was used to determine test-retest reliability of the new protocol of measuring hand volume with the commercial volumeter. The new protocol used a height-adjustable table and external trunk support of the participant in order to maintain consistency of web-space pressure between the first and second measurement. A 1 ml micropipette was used to obtain a more accurate volume reading. An intraclass correlation (2,1), repeated measures ANOVA, and standard error of measurement were used to determine the results of this study.
Results:
This study confirmed the test-retest reliability of the commercial volumeter and decreased the standard error of measurement to 3~ml as opposed to the current standard of 10 ml.
Conclusion:
Clinically, this study is significant in that occupational and physical therapists can detect more subtle change in hand volume.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Uncertainty in Measurement: Reading Instruments
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Assessment of blood pressure in brachial artery(two-step method)
Pipe Flowrate Measurement
The orifice meter is a simple,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...