Aim:
To study the effectiveness of an electronic apex locator (EAL; Justy II; Yoshida Dentcraft, Tokyo, Japan) in locating simulated horizontal and vertical fractures in single roots.
Methodology:
An EAL was used to measure the distance within the canal of horizontal (n = 31) and vertical (n = 31) fractures, created with a disk in single-rooted teeth. Accuracy of the EAL was evaluated by comparing the measurements with those made using a size 10 file. Data were analysed with the non-parametric Passing and Bablok method.
Results:
For simulated horizontal fractures, the EAL measured exactly the same length as a size 10 file, without constant or proportional errors. In vertical simulated fractures, the EAL measured (on average) with a constant error of 7.5 mm shorter than the size 10 file; the difference had a wide confidence interval (-72.3 to 2.6 mm).
Conclusions:
In this laboratory study, the Justy II EAL was able to determine accurately the position of simulated horizontal fractures, but was unreliable when measuring simulated vertical fractures.
Related Concept Videos
Adjusting a Traverse
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...
Angle of Twist - Elastic Range
Stresses in a Shaft
Applying equilibrium conditions to the QR segment establishes that the internal shearing forces within the...
Common Leveling Mistakes and Errors