Endotracheal suctioning: from principles to practice
Brenda M Morrow1, Merle J Futter, Andrew C Argent
1Physiotherapy Department, Red Cross War Memorial Children's Hospital, School of Health and Rehabilitation Sciences, University of Cape Town, Klipfontein Road, 7700 Rondebosch, Cape Town, South Africa. bmorrow@ich.uct.ac.za
Insights
Endotracheal suctioning in neonates can cause significant pressure changes, potentially leading to lung volume loss. Optimizing catheter size and suction pressure is crucial for safe paediatric practice.
Area of Science:
- Pediatric Intensive Care
- Respiratory Physiology
- Medical Device Engineering
Background:
- Endotracheal suctioning is a common procedure in neonatal intensive care.
- Understanding the physical principles governing suctioning is essential for patient safety.
- Previous studies have not fully elucidated the pressure dynamics involved in neonatal endotracheal suctioning.
Purpose of the Study:
- To investigate the physical principles of pressure and flow dynamics during endotracheal suctioning.
- To model these principles using an in vitro system.
- To discuss the clinical implications for pediatric practice.
Main Methods:
- Prospective in vitro study conducted in a pediatric intensive care unit.
- Utilized a "bag-in-box" model to simulate neonatal lungs.
- Measured peak pressure changes (DeltaP) with varying endotracheal tube (ETT) and catheter sizes, suction pressures, and techniques.
- Assessed suctioning of different gelatin consistencies to quantify mucus removal.
Main Results:
- A large DeltaP was observed with neonatal-sized ETTs.
- DeltaP showed a linear relationship with the catheter-to-ETT area ratio.
- Shorter catheters, longer suction duration, and higher vacuum pressures resulted in significantly greater DeltaP.
- Mucus suctioned per unit time correlated with catheter size, suction pressure, and mucus density.
Conclusions:
- Endotracheal suctioning in intubated neonates generates considerable intrapulmonary pressure changes.
- These pressure changes may lead to loss of lung volume.
- Findings emphasize the need for careful selection of suction parameters in pediatric practice.
Objective:
This paper aims to highlight the physical principles of pressure and flow dynamics underlying endotracheal suctioning, to demonstrate these on a simple illustrative model and to discuss the clinical implications of these principles for paediatric practice.
Design:
Prospective in vitro study.
Setting:
Paediatric intensive care unit of a tertiary, academic hospital.
Measurements And Main Results:
The peak pressure change (DeltaP) obtained in a "bag-in-box" model was recorded during suctioning, using different combinations of endotracheal tube (ETT) and catheter sizes, different suction pressures and techniques. Suction was also performed on three different consistencies of gelatine, using a range of suction catheter sizes and vacuum pressures. The mass of gelatine suctioned per second was calculated. A large DeltaP was measured when using neonatal-sized ETTs. There was a linear relationship (r=0,8, p<0.05) between DeltaP and the ratio of external catheter area to area difference (internal ETT area-external catheter area). Significantly greater DeltaP was measured when using a short versus long suction catheter ( p<0.001) and when applying suction for longer duration ( p<0,001) and with higher vacuum pressures ( p<0.05). The amount of mucus suctioned in a given time was related to catheter size, suction pressure and mucus density.
Conclusion:
Applied clinically, these results indicate that intrapulmonary pressure changes generated by the endotracheal suctioning of intubated neonates are likely to be considerable, possibly translating into loss of lung volume.
Related Concept Videos
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Tracheostomy Suctioning II: Procedure
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.